Fct.

Text

Description and settings

3.03

ZERO SET

Zero calibration
Note:

Carry out only at “0” flow and with completely filled
measuring tube!

Precautionary query
CALIB. NO

Press

key to return to Fct. 3.03 “ZERO SET”.

CALIB. YES Press

key to start calibration.

Duration approx. 15-90 seconds, (depends on frequency
of magnetic field) current flowrate displayed in the selected
unit (see Fct. 1.04 “DISP. FLOW”).

A “WARNING” sign appears when flowrate “> 0“;
acknowledge by pressing

key.

STORE NO (do

not store new zero value)

STORE YES (store new zero value)
Press

key to return to Fct. 3.03 “ZERO SET”.

3.04

ENTRY CODE

Entry code required to enter setting mode?
NO

(= entry with

→ only)

YES (= entry with

→ and Code 1: → → → ↵ ↵ ↵ ↑ ↑ ↑ )

Press

to return to Fct. 3.04 “ENTRY CODE”.

3.05

USER UNIT

Set any required unit for flowrate and counting

TEXT VOL.

Set text for required flowrate unit (max. 5 characters)
Factory-set: “Liter” or “MGal”.
Characters assignable to each place:
A-Z, a-z, 0-9, or “ – “ (= blank character).
Press

key to transfer to subfunction “FACT. VOL.”

FACT. VOL.

Set conversion factor (F

M

) for volume

Factory set “1.00000 E+3” for “Liter” or “2.64172E-4” for “US MGal”
(exponent notation, here: 1

x

10

3

or 2.64172

x

10

-4

).

Factor F

M

= volume per 1m

3

.

Setting range

1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

key to transfer to subfunction “TEXT TIME”.

TEXT TIME

Set text for required time unit (max. 3 characters)
Factory-set: “hr” (=hour).
Characters assignable to each place:
A-Z, a-z, 0-9, or “ – “ (= blank character).
Press

↵ key to transfer to subfunction “FACT. TIME”

FACT. TIME

Set conversion factor (F

T

) for time

Factory-set: “3.60000 E+3” for “hour” or “8.64000 E+4” for “day”
(exponent notation, here: 3.6

x

10

3

or 8.64

x

10

-4

).

Set factor F

T

in seconds.

Setting range
1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

key to return to Fct. 3.05 “USER UNIT”.

3.06

APPLICAT.

Set overload point for A/D converter

FLOW

STEADY (150% of Q

100%

) PULSATING (1000% of Q

100%

)

Press key

to return to Fct. 3.06 “APPLICAT.“,

3.07

MEASURING POINT

Set measuring point no.
Factory setting: ALTOMETER
Characters assignable to each place:
A-Z, a-z, 0-9, or “ – “ (= blank character).
Press

↵ key to return to Fct. 3.07”MEASURING POINT”

3.08

COM

Set comminucation interface
OFF (switched off)
HART (HART-interface switched on)
KROHNE (RS 485 interface switched on)
Address:”HART” 00-15 / “KROHNE” 000-239
BAUD RATE: -1200

-2400

-4800

-9600

-19200

(appeare with selection “KROHNE”)

Press key

to return to Fct. 3.08 “COM”.

4/5