KROHNE IFC 020 Short EN User Manual

Page 17

background image

Condensed instructions IFC 020

4/3

Fct.

Text

Description and settings

1.05

CURRENT I

Current output I

FUNCT. I

Select function for current output I
OFF (switched off)
1 DIR. (1 flow direction)
2 DIR. (forward/reverse flow, F/R flow measurement)
Press key

, transfer to subfunction “RANGE I”;

RANGE I

Select measuring range
0 - 20 mA 4 - 20 mA (fixed ranges)
mA (user-defined range) I

0%

- I

100%

(Value I

0%

< I

100%

!) 0 - 16 mA 4 - 20 mA

To transfer to number setting, press key

!

Press key

to transfer to subfunction “I ERROR“.

I ERROR

Select error value
22 mA 0.0 to I

0%

mA (variable, see above if I

0%

> 1 mA)

To transfer to number setting, press key

!

Press key

to revert to Fct. 1.05 CURRENT. I.

1.06

PULS B1

Pulse output B1 (see Fct. 3.07 HARDWARE)

FUNCT. P

Select function for pulse output P
OFF (switched off)
1 DIR. (1 flow direction)
2 DIR. (forward/reverse flow, F/R flow measurement)
Press key

to transfer to subfunction “SELECT P”.

SELECT P

Select pulse type
100 HZ

PULSE/VOL. (pulses per unit volume, flowrate)

1000 HZ

PULSE/TIME (pulses per unit time for 100% flowrate)

Press key

to transfer to subfunction “PULSWIDTH”.

With selection 100 HZ and 1000 HZ,
return to Fct.1.06 pulsoutput P (pulswidth 50% cyclic).

PULSWIDTH

Select pulse width
50 msec 100 msec 200 msec 500 msec 1 sec
Press key

to transfer to subfunction “VALUE P”.

VALUE P

Set pulse value per unit volume (appears only when
“PULSE/VOL.” set under “SELECT P” above)
xxxx PulS/m3 xxxx PulS/Liter xxxx PulS/US.Gal
xxxx PulS/ user-defined unit, factory-set is “Liter” or “US M.Gal” (see Fct. 3.05)
Setting range “xxxx” is dependent on the pulse width and the
full-scale range: P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Press key

to return to Fct.1.06 “PULS. OUTPUT P”.

VALUE P

Set pulse value per unit time (appears only when “PULSE/TIME”
set under “SELECT P” above)
xxxx PulSe/Sec (=Hz) xxxx PulSe/min xxxx PulSe/hr
xxxx PulSe/user-defined unit, factory-set is “hr” (see Fct. 3.05)
Setting range “xxxx” is dependent on the pulse width, see above
Press key

to return to Fct.1.06 “PULS. OUTPUT P””.

1.07

STATUS. S

Status output S

ALL ERROR

FATAL ERROR

OFF

ON

F/R INDIC. (F/R indication for forward/reverse measurement)
TRIP. POINT

Setting range: 002 - 115 PERCENT

EMPTY PIPE (appears only when this option is installed)
Press

key to transfer to number setting

Press

key to return to Fct. 1.07 “STATUS. S“.

1.08

CONTROL E

Control input E (IFC 020 E only)

(IFC 020 E only)

OFF (switched off)
OUTP. ZERO (set outputs to “min.values”)
TOTAL. RESET (reset totalizers)
ERROR. RESET(delete error messages)
Press

key to return Fct. 1.08 “Control E”.