Condensed instructions IFC 020

4/4

Fct.

Text

Description and settings

2.00

TEST

Testmenu

2.01

TEST Q

Test measuring range Q
Precautionary query
SURE NO

Press

key to return to Fct. 2.01 “TEST Q”.

SURE YES

Press

key, then use key to

select value: -110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.

of set full-scale range Q

100%

.

Displayed value present at outputs I and P.

Press

key to return to Fct. 2.01 “TEST Q”.

2.02

HARDW. INFO

Hardware information and error status
Before consulting factory, please note down all 6 codes.

MODUL ADC

X . X X X X X . X X

Y Y Y Y Y Y Y Y Y Y

Press

key to transfer to “MODUL IO“.

MODUL IO

X . X X X X X . X X

Y Y Y Y Y Y Y Y Y Y

Press

key to transfer to “MODUL DISP.“.

MODUL DISP.

X . X X X X X . X X

Press

key to transfer to “MODUL RS

Y Y Y Y Y Y Y Y Y Y

MODUL RS

X . X X X X X . X X

Press

key to return to Fct. 2.02

Y Y Y Y Y Y Y Y Y Y

“HARDW. INFO“.

Fct.

Text

Description and settings

3.00

INSTALL.

Installation menu

3.01

LANGUAGE

Select language for display texts
GB / USA (English)

F (French)

D (German)

others on request

Press

key to return to Fct. 3.01 “LANGUAGE”.

3.02

FLOWMETER

Set data for primary head

DIAMETER

Select size from meter size table
DN 2.5 - 1000 mm equivalent to

1

/

10

- 40 inch

Select with

key.

Press

key to transfer to subfunction “FULL SCALE”.

FULL SCALE

Full-scale range for flow Q

100%

To set, refer to Fct. 1.01 “FULL SCALE” above.
Press

↵ key to transfer to subfunction “GK VALUE”.

VALUE P

Pulse value (Fct. 1.06 “VALUE P”) has been changed.

With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Check new values!

GK VALUE

Primary head

- Set primary constant GK

see primary head nameplate.
Range:

1.0000 - 9.9999

Press

key to transfer to subfunction “FIELD. FREQ.”.

FIELD FREQ.

Magnetic field frequency
Values: 1/6 and 1/18 of power frequency, see nameplate.
Press

key to transfer to subfunction “FLOW DIR.”;

DC units only: to transfer to subfunction “LINE FREQ.”.

LINE FREQ.

Normal line frequency in your country
Please note: This function is only provided for units
with DC power supply to suppress line-frequency interference.
Values: 50 Hz and 60 Hz
Press

key to transfer to subfunction “FLOW DIR.”.

FLOW DIR.

Define flow direction (in F/R mode: forward flow).
Set according to direction of arrow on primary head:
+ DIR.

– DIR. Select using

key.

Press

key to return to Fct. 3.02 “FLOWMETER”.