2/4

Condensed instructions IFC 020

U

ext

≤ 30 V DC/≤ 24 V AC

I

max

≤ 6 mA

8

7

Control input E

passive

(IFC 020 E only)

RS 485 Interface

For connection and
operation of the
Krohne RS 485
Interface,
see section 6.2.

6

5

Status output S

passive

U

ext

≤ 30 V DC/≤ 24 V AC

I

max

≤ 150 mA

(incl. pulse output P)

Status output S

active

with and without current output I

U

int

≤ 15 V DC

from current output

I

max

≤ 3 mA

Operation with
current output

I

max

≤ 23 mA

Operation without
current output

Load

Load

Error messages in measuring mode 2.5

The following list gives all errors that can occur during process flow measurement.
Errors shown in display when “YES“ set in Fct. 1.04 DISPLAY, subfunction “DISP. MSG.“.

Error messages

Description of error

Error clearance

LINE INT.

Power failure Note:

Cancel error in RESET/QUIT. menu

no counting during power failure

Reset totalizer if necessary.

CUR. OUTP. I

Current output overranged.

Check and if necessary correct

(Flow > measuring rage)

instrument parameters. After
elimination of cause, error message
deleted automatically.

PULSOUTP. P

Pulse output overranged.

Check and if necessary correct

Note:

instrument parameters. After

totalizer deviation possible.

elimination of cause, error message
deleted automatically.

TOTALIZER

Totalizer has been reset

Cancel error message in
RESET/QUIT. menu, see section 4.6.

ADC

Analog / digital

Error message deleted automatically

converter overranged

after elimination of cause.

FATAL. ERROR

Fatal error, all outputs set

Please consult factory.

to “min. values“