beautypg.com

3 instrument repair – KROHNE SU 600 Ex EN User Manual

Page 24

background image

?

E016

l

Empty/full adjustment reversed

à

Carry out a fresh adjustment

?

E017

l

Adjustment span too small

à

Carry out a fresh adjustment and increase the distance

between min. and max. adjustment

?

E021

l

Scaling span too small

à

Carry out a fresh scaling, increase the distance

between min. and max. scaling.

?

E110

l

Relay switching points too close together

à

Increase the difference between the two relay switching

points

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form from our Internet homepage

http://www.krohne-mar.com/fileadmin/media-lounge/PDF-

Download/Specimen_e.pdf.
By doing this you help us carry out the repair quickly and

without having to call back for needed information.

l

Print and fill out one form per instrument

l

Clean the instrument and pack it damage-proof

l

Attach the completed form and possibly also a safety data

sheet to the instrument.

24

SU 600 Ex - 4 ... 20 mA signal conditioning

Maintenance and fault rectification

30653

-EN
-050803