7 Maintenance and fault rectification

7.1 Maintenance

When used as directed in normal operation, SU 600 Ex is

completely maintenance-free.

7.2 Fault rectification

SU 600 Ex offers maximum reliability. Nevertheless faults can

occur during operation. These may be caused by the following,

e.g.:

l

Measured value of the sensor not correct

l

Power supply

l

Interference on the cables

The first measures to be taken are to check the input/output

signals as well as to evaluate the error messages via the

display. The procedure is described below. In many cases, the

causes can be determined in this way and faults can be

rectified.

?

E003

l

CRC-error

à

Carry out a reset

à

Return instrument for repair

?

E014

l

Short-circuit

à

Eliminate short-circuit

à

If necessary, exchange defective sensor

?

E015

l

Line break

à

Repair sensor cable or replace it

à

Check sensor connection