Level

2

Mode overfill protection

3

Mode dry run protection

Scaling is the conversion of the measured value into a certain

parameter or measuring unit. Instead of the percentage value,

the volume can be displayed, e.g. in l. Indicating values

between max. -9999 and +9999 are possible.

à

First of all, determine the max. indicating range and the

number of decimal places after the decimal point -- max.

four decimal places can be displayed. Select position [ 5 ]

of the function switch, place the decimal point in the

required position and save your settings.

à

Now select position [6] or [7] on the function switch, enter

the requested values for 100 % or 0 % and save your

settings.

To suppress fluctuation in the measured value display, e.g.

caused by an agitated product surface, an integration time can

be set. This time can be between 0 and 250 seconds. Keep in

mind that this setting will increase the reaction time of the

measurement, and that the reaction to quick changes of the

measured value will be delayed. In general, a time of a few

seconds is sufficient to smooth the measured value display.

Scaling

Damping

SU 600 Ex - 4 ... 20 mA signal conditioning

19

Setup with the integrated indicating and adjustment unit