beautypg.com

Technical data, 2 technical data for the sensor – KROHNE OPTISENS CL 1100 EN User Manual

Page 43

background image

TECHNICAL DATA

7

43

OPTISENS CL 1100

www.krohne.com

01/2012 - 4001834501 - MA OPTISENS CL 1100 R01 en

7.2 Technical data for the sensor

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

INFORMATION!
For further information about the technical data please consider the MAC 100 converter manual.

OPTISENS CL 1100

Cl

2

ClO

2

O

3

Measuring system

Measuring principle

Potentiostatic with double gold electrodes

Application range

Continuous

measurement of free

chlorine in water

applications.

Continuous

measurement of

chlorine dioxide in water

applications.

Continuous

measurement of ozon in

water applications.

Measuring range

Cl

2

: 0…5 mg/l

ClO

2

: 0…5 mg/l

O

3

: 0…5 mg/l

Design

Construction

Glass electrode

Shaft diameter

12 mm / 0.47"

Lenght

120 mm / 4.72"

Process connection

PG 13.5

Sensor cap

M12

Measuring accuracy

Reference conditions

Medium: water

Temperature: 20°C / 68°F

Pressure: 1 bar / 14.5 psi (absolute)

Maximum measuring

error

≤ 2% full scale
Temperature: 1.0% full scale

Repeatability

0.2% full scale

Resolution

0.1/0.01 mg/l

Long-term stability

24 hours: tested within accuracy definition

Temperature drift

Tested within accuracy definition

Cable length variation

Tested within accuracy definition

.book Page 43 Wednesday, February 1, 2012 11:58 AM