5

OPERATION

26

OPTISENS CL 1100

www.krohne.com

01/2012 - 4001834501 - MA OPTISENS CL 1100 R01 en

Step 2: Configure/adjust the temperature sensor for measurement

Press ^

^

^

^ to confirm the entered value. The temperature sensor has been adjusted.

If necessary, enter the temperature correction in Kelvin so that the signal converter shows the
same temperature as the reference thermometer.

Step 2a: probe Pt100

probe Pt100

probe Pt100

probe Pt100 or Pt1000:

Pt1000:

Pt1000:

Pt1000:

Read the currently measured temperature of the Pt100 / Pt1000 temperature sensor from the

measurement screen and write it down.

Measure the temperature with a reference thermometer and check if it deviates from the temperature

measured by the Pt100 / Pt1000.

Press >>>> for more than 2.5 seconds, then release the button. You are now on the main menu level.

Press

or

until the main menu setup

setup

setup

setup is highlighted.

MAIN MENU

A quick setup

B test

> C setup

> C setup

> C setup

> C setup

D service

Press >>>> to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "SETUP"

and "C1" appears, beneath the submenu process input A

process input A

process input A

process input A is highlighted.

Press >

>>> to enter the chosen menu process input A

process input A

process input A

process input A.

Press

or

until the main menu temperature

temperature

temperature

temperature is highlighted.

Press >

>>> to enter the chosen menu. The submenu probe Pt100 / 1000

probe Pt100 / 1000

probe Pt100 / 1000

probe Pt100 / 1000 is

highlighted.

Press

or

until the submenu correction

correction

correction

correction is highlighted.

Press >

>>> to enter the chosen menu.

INFORMATION!
A separate temperature sensor has to be connected to "Pos.A" and configurated on process
input A.