TECHNICAL DATA

44

OPTISENS PH 8100

www.krohne.com

03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en

The voltage change of a pH electrode at 25°C / 77°F is around 59 mV for each pH unit. This is also
called the slope of the pH electrode. The slope is temperature dependent and decreases over life
time of the electrode.

To compensate for the temperature dependency of the pH measurement, the temperature of the
medium can be measured and automatically compensated in the signal converter.

Figure 6-2: Optimal slope of the electrode at 25°C / 77°F

Figure 6-3: Temperature dependency of the electrode slope