Operation – KROHNE OPTISENS PH 8100 EN User Manual

Page 34

background image

4

OPERATION

34

OPTISENS PH 8100

www.krohne.com

03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en

Step 1: activating the manual hold function

• You have activated the manual hold function. To go to the next step and prepare the calibration

procedure. You have to return to the measuring mode.

• Press ^ until you reach the measuring mode again

Step 2: preparing the calibration procedure

• If you re-calibrate an existing electrode, remove the electrode from its respective assembly

(for further information refer to

Mounting to a flow-through holder on page 18 or refer to

Calibrating pH measurement on page 33).

• If you calibrate a new electrode, make sure that the electrode is correctly connected to the

converter.

• Check the electrode for damages, check the diaphragm for coating and rinse the electrode tip

with tap water and gently swipe it with a soft tissue.

• Provide two buffer solutions with the same known temperature: pH 4 and pH 7.

After activating the manual hold function and the preparative measures, you can get access to
the calibration procedure from the measuring mode in two different ways. Either you go via the
main menu setup

setup

setup

setup (step 3a) or via the main menu quick setup

quick setup

quick setup

quick setup (step 3b).

Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup

quick setup

quick setup

quick setup is highlighted.

Press

or

until the main menu quick setup

quick setup

quick setup

quick setup is highlighted.

MAIN MENU

> A quick setup

> A quick setup

> A quick setup

> A quick setup

B test

C setup

D service

Press >

>>> to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "quick setup" and

A1" appears, beneath the submenu language

language

language

language is highlighted.

Press

or

until the submenu manual hold

manual hold

manual hold

manual hold is highlighted.

Press >

>>> to enter the chosen menu.

You are on the second submenu level. In the upper line of the display

"manual hold" appears, beneath the option off

off

off

off is highlighted

Press

or

to choose the option on

Press ^ to confirm the entered value.

.book Page 34 Wednesday, July 4, 2012 11:07 AM