Operation – KROHNE OPTISENS PH 8100 EN User Manual

Page 28

background image

4

OPERATION

28

OPTISENS PH 8100

www.krohne.com

03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en

C1.15.5

C2.15.5

temp. comp.

Menu for activating the temperature compensation parameters for the

measurement.

Options:

Linear: linear temperature compensation.

Off: temperature compensation is disabled.

C1.16

C1.16

C1.16

C1.16

C2.16

C2.16

C2.16

C2.16

ORP cal.

menu item for calibrating the ORP sensor

C1.16.1

C2.16.1

prepare calibration

view actual ORP value

C1.16.2

C2.16.2

ref. value ORP

Enter the reference value of the probe in mV

C1.16.3

C2.16.3

start calibration?

Start calibration procedure.

C1.16.4

C2.16.4

stored value

View stored value of calibration.

C1.17

C1.17

C1.17

C1.17

C2.17

C2.17

C2.17

C2.17

pH cal.

menu item for calibration the pH sensor

C1.17.1

C2.17.1

temp.comp

Menu for activating the temperature compensation parameters for the

calibration

Options:

off: temperature measurement is disabled

manual: temperature value has to be entered manually

automatic: temperature measurement is performed as configured

C1.17.3

C2.17.3

calib. buffer 1

Start calibration procedure in buffer 1

C1.17.4

C2.17.4

pH 1 set value

Enter the reference value of the probe in pH

C1.17.6

C2.17.6

calib. buffer 2

Start calibration procedure in buffer 2

C1.17.7

C2.17.7

pH 2 set value

Enter the reference value of the probe in pH

CAUTION!
If you choose for measurement the temperature compensation "linear" than choose between
"automatic" or "manual" for the temperature compensation during calibration. If you choose for
measurement the temperature compensation "off" than choose also "off" for the temperature
compensation during calibration.

Level

Designation / function

Settings / descriptions

.book Page 28 Wednesday, July 4, 2012 11:07 AM