Installation, 6 calibrating the electrode – KROHNE OPTISENS PH 8100 EN User Manual

Page 16

background image

3

INSTALLATION

16

OPTISENS PH 8100

www.krohne.com

03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en

3.5.3 Connecting the cable to the sensor

Connecting the cable to the sensor

• Unscrew the protective cap from the sensor connector and keep it for future use 1.
• Ensure that both cable and electrode connector are absolutely dry 2.
• Make sure that the O-ring is positioned on the sensor connector 3.
• Screw the cable connector 4 on to the sensor and tighten it by hand.

3.6 Calibrating the electrode

Before the electrode is installed, it has to be calibrated. Proceed as described on page 29. Then
continue with the installation procedure.

CAUTION!
Moisture inside the sensor connector must be avoided! Moisture will shortcut the mV signal
between the electrodes and deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).

Figure 3-6: Connecting the cable to the sensor