INSTALLATION

3

9

PTD 600

www.krohne.com

03/2012 - 4001946801 - MA PTD 600 R01 en

Installation

3.1 Notes on installation

3.2 Installation specifications

Even though its design is robust, the pressure transmitter should not be exposed to any heavy
impact. Avoid high static and dynamic pressures that exceed the values indicated in "Technical
data".

In the case of flush connections, there is one wavy, highly sensitive membrane exposed. To avoid
damaging the membrane, do not use abrasive cleaning agents or allow it to come into contact
with hard bodies (including fingers) or tools. The protective cover should thus only be removed
immediately before installing the transmitter.

3.3 Mounting to pressure system

Versions with thread connection

Versions with thread connection

Versions with thread connection

Versions with thread connection

• The pressure transmitter features a male thread for the connection.
• If the seal is metallic, it is recommended to grease the raised face with a MoS

2

or Vaseline-

based lubricant.

• Use a suitable wrench for mounting. Do not exceed the tightening torque when mounting

(refer to the table below).

Versions with hygienic clamp connection

Versions with hygienic clamp connection

Versions with hygienic clamp connection

Versions with hygienic clamp connection

• The pressure transmitter features appropriate hygienic clamp connections for the

connection.

• Only use clamp connections designed for the specific application and pressure.
• Do not damage the seal during installation or replacement.

INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.

INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.

INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.

Process
connection

Tightening torques [Nm] at a pressure of

≤ 1 bar / 14.5 psi

1 bar / 14.5 psi > p ≤40 bar / 580 psi

10

15

G1

40