beautypg.com

Technical data, Technical data 7.1 technical data – KROHNE PTD 600 EN User Manual

Page 17

background image

TECHNICAL DATA

7

17

PTD 600

www.krohne.com

03/2012 - 4001946801 - MA PTD 600 R01 en

Technical data

7.1 Technical data

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Measuring principle

Piezoresistive silicon sensor

Application range

Hydrostatic level detection as well as pressure measurement in

pipelines for gas, steam and liquids

Measuring range

-1...40 bar / -14.5...580 psi

Measuring accuracy

Reference conditions

Medium: air

Temperature: +20°C / +68°F

Pressure type

Gauge pressure / absolute pressure

Measuring accuracy

0.5%, 0.25% or 0.1% of the full scale (including linearity, hysteresis

and repeatability)

Deviation of zero point

≤±0.03% of the full scale/10 K

Deviation of span

≤±0.03% of the full scale/10 K

Long-term stability

0.1% of the full scale/ year

Response time (10...90%)

5 ms

Operating conditions

Temperature

Temperature

Temperature

Temperature

Process temperature

Standard: -10…+125°C / +14...+257°F

With optional cooling neck: -10…+200°C / +14...+392°F

CIP/SIP

Max. up to 150°C / 302°F < 60 minutes for hygienic process

connections

Max. up to 125°C / 257°F for devices with thread connection

Ambient temperature

-10…+85°C / +14...+185°F

Storage temperature

-10…+85°C / +14...+185°F

Other conditions

Other conditions

Other conditions

Other conditions

Protection category acc. to IEC 529 / EN 60529

IP67

Installation conditions

Installation

Can be installed in any position, zero point or position correction may

be required following installation.

To ensure proper cooling function, versions with cooling neck

(T

Medium

≤ 200°C / 392°F) must be installed at a maximum 10° angle

to the horizontal.

Dimensions and weights

For detailed information refer to section "Dimensions and weights".

.book Page 17 Friday, March 23, 2012 10:14 AM