beautypg.com

Assembled equipment – KROHNE BM 26 BASIC-ADVANCED ATEX EN User Manual

Page 25

background image

ASSEMBLED EQUIPMENT

4

25

BM 26 BASIC/ADVANCED

www.krohne.com

12/2011 - 4000347102 - AD ATEX BM 26 Bas-Adv R02 en

Maximum ambient temperature: measuring chamber with an analog transmitter (FF and
PROFIBUS PA outputs) - Part 2

Temperature class

Maximum ambient temperature

FOUNDATION™ Fieldbus (xF45xD1xx) / PROFIBUS PA (xF45xX1xx)

- PR

PROFIBU

S PA

(xF45xA1

xx)

Conventional

intrinsically-safe

systems

FISCO intrinsically-safe systems

Alternative

1

2

3

4

5

6

T6

[°C]

+59

+44

+44

+44

+59

+59

+44

[°F]

+138

+111

+111

+111

+138

+138

+111

T5

[°C]

+69

64

+59

+59

+73

+73

+59

[°F]

+156

+147

+138

+138

+163

+163

+138

T4

[°C]

+80

+74

+80

+80

+80

+80

+80

[°F]

+176

+165

+176

+176

+176

+176

+176

T3

[°C]

+80

1

+73

1

+80

1

+80

1

+80

1

+80

1

+80

1

[°F]

+176

1

+163

1

+176

1

+176

1

+176

1

+176

1

+176

1

T2

[°C]

+80

1

+72

1

+80

1

+80

1

+80

1

+80

1

+80

1

[°F]

+176

1

+162

1

+176

1

+176

1

+176

1

+176

1

+176

1

T1

[°C]

+74

1

+74

1

+74

1

+74

1

+74

1

+74

1

+74

1

[°F]

+165

1

+165

1

+165

1

+165

1

+165

1

+165

1

+165

1

1 The analog transmitter is certified for applications with temperature classes T6...T4. If the ambient temperature for this device is not

more than the maximum limit for temperature classes T6...T4, the analog transmitter can be attached to a measuring chamber for
applications with temperature classes T6...T1.

INFORMATION!
Measuring Chamber equipped with an analog transmitter and limit switches

Measuring Chamber equipped with an analog transmitter and limit switches

Measuring Chamber equipped with an analog transmitter and limit switches

Measuring Chamber equipped with an analog transmitter and limit switches
The analog transmitter and limit switches are certified for applications with temperature classes
T6...T4. If the ambient temperature for these devices is not more than the maximum limit for
temperature classes T6...T4, the analog transmitter and limit switches can be attached to a
measuring chamber for applications with temperature classes T6...T1.
Use the lowest value in the tables in this section to get the maximum temperature limit for the
measuring chamber with analog transmitter and limit switch options.