A minimum operating pressure (primary pressure) is necessary to operate the device:

CAUTION!

When starting up the device, the following points must be observed:

Compare the actual operating pressure and the product temperature of the system with the
specifications on the nameplate (PS and TS). These specifications may not be exceeded.

Make sure materials are compatible.

Slowly open the shut-off valve.

When measuring liquids, vent the pipes carefully.

When measuring gases, increase pressure slowly.

Avoid float impact (e.g. caused by solenoid valves), as this is likely to damage the measuring
unit or float.

Medium

Pressure loss : operating pressure

Liquids

1 : 2

Gases without float damping

1 : 5

Gases with float damping

1 : 2