4.3 Ground wire

4.4 Ingress protection

The measuring device meets all requirements of protection category IP66/67.

Therefore it is essential to observe the following points:

• Use only original gaskets. They must be clean and free of any damage. Defective gaskets must

be replaced.

• The electrical cables used must be undamaged and must comply with regulations.
• The cables must be laid with a loop 3

3

3

3 upstream of the measuring device to prevent water

from getting into the housing.

• The cable feedthroughs 2

2

2

2 must be tightened.

• Close the unused cable feedthroughs using blind plugs 1

1

1

1.

DANGER!
The grounding wire may not transfer any interference voltage.
Do not use this grounding cable to ground any other electrical devices.

DANGER!
After all servicing and maintenance work on the measuring device, the specified protection class
must be ensured again.

Figure 4-1: Cable feedthrough

1 Use blanking plugs if no cable is routed through
2 Tighten cable feedthrough firmly
3 Lay the cable in a loop