Maintenance and service, 3 replacement of mains fuse – KROHNE OPTIFLUX 2000-4000-5000-6000-7300-IFC 300 Ex EN User Manual

Page 38

background image

5

MAINTENANCE AND SERVICE

38

OPTIFLUX

www.krohne.com

07/2010 - 7309482600 - AD EX IFC 300 R06 en

5.3 Replacement of mains fuse

• Pull the display unit of the mounting frame using the two metal levers left and right and turn

display unit carefully aside.

• Unscrew the two screws size M4 that hold the mounting frame with the electronics unit.
• Carefully pull the mounting frame with electronics unit almost completely out of the housing,

disconnect the long rectangular (14-pole) blue connector at the back-end of the electronics
unit. Now carefully remove the unit from the housing.

• The mains fuse is located in a fuse holder at the back-end of the electronics unit. The

specifications must be as follows:

• Reassemble the unit in reverse order.

WARNING!
Before commencing the work, refer to Before and after opening on page 37

.

Fuse type: 5 x 20 mm (H) according to IEC 60127

Fuse type: 5 x 20 mm (H) according to IEC 60127

Fuse type: 5 x 20 mm (H) according to IEC 60127

Fuse type: 5 x 20 mm (H) according to IEC 60127

Power supply

Power supply

Power supply

Power supply

Electrical data

Electrical data

Electrical data

Electrical data

12...24 V DC

250 V / 2 A

24 V AC/DC

250 V / 2 A

100...230 V AC

250 V / 1.6 A

WARNING!
Before reassembling the unit, refer to Before and after opening on page 37

.