beautypg.com

Temperature limits, 1 compact versions – KROHNE OPTIFLUX 2000-4000-5000-6000-7300-IFC 300 Ex EN User Manual

Page 22

background image

2

TEMPERATURE LIMITS

22

OPTIFLUX

www.krohne.com

07/2010 - 7309482600 - AD EX IFC 300 R06 en

2.2 OPTIFLUX 5000

The maximum process temperature T

p

is dependent on the required temperature class T6...T3

and the maximum ambient temperature T

a

.

2.2.1 Compact versions

Sandwich versions

Sandwich versions

Sandwich versions

Sandwich versions

• For dust hazardous areas, the maximum surface temperature is equal to the process

temperature with a minimum of 85 °C.

• The OPTIFLUX 5300 C "sandwich" is suitable for an ambient temperature range of

-20...60°C (DN2.5...15) or -40...+60°C (DN25...100).

• The minimum process temperature is -20°C (DN2.5...15) or -40°C (DN25...100).

Flange versions

Flange versions

Flange versions

Flange versions

• The OPTIFLUX 5300 C "flange" is suitable for an ambient temperature range of

-40...+60°C.

• The minimum process temperature is -40°C.

Temperature class

Max. process temperature Tp [°C]

T

a

≤ 40°C

40 < T

a

≤ 50°C

50 < T

a

≤ 60°C

T6

60

55

-

T5

75

75

70

T4

115

115

75

T3

150

135

75

Temperature
class

Max. surface
temperature for
dust [°C]

Max. process temperature Tp [°C]

T

a

≤ 40°C

40 < T

a

≤ 50°C

50 < T

a

≤ 60°C

T6

85

80

80

60

T5

100

95

95

60

T4

135

135

130

60

T3

150

150

145

60