CONNECTION DIAGRAMS FOR FIELD VERSIONS

30

OPTIFLUX

www.krohne.com

07/2010 - 7309482600 - AD EX IFC 300 R06 en

3.4 Signal cable connections

Figure 3-3: Connecting with signal cable A

1 Equipotential bonding, conductor ≥ 4 mm

2

2 Field current cable acc. EN 60079-14 clause 9.3 and 11.3 (Increased safety), do not connect the screen if available.
3 Signal cable A acc. EN 60079-14 clause 12.2 (Intrinsic safety)

CAUTION!
* Outer screen connected via strain reliefs.

Figure 3-4: Connecting with signal cable B

1 Equipotential bonding, conductor ≥ 4 mm

2

2 Field current cable acc. EN 60079-14 clause 9.3 and 11.3 (Increased safety), do not connect the screen if available.
3 Signal cable B acc. EN 60079-14 clause 12.2 (Intrinsic safety)

CAUTION!
* Outer screen connected via strain reliefs.