Connection diagrams for field versions, 2 signal cable b – KROHNE OPTIFLUX 2000-4000-5000-6000-7300-IFC 300 Ex EN User Manual

Page 28

background image

3

CONNECTION DIAGRAMS FOR FIELD VERSIONS

28

OPTIFLUX

www.krohne.com

07/2010 - 7309482600 - AD EX IFC 300 R06 en

3.2 Signal cable B

The signal cable B is a triple screen shielding cable, according to EN 60079-14 clause 12.2
(Intrinsic safety).

Figure 3-2: Construction of signal cable B

1 Stranded drain wire for the inner shield (10), 1.0 mm

2

Cu / AWG 17 (not insulated, bare)

2 Insulated wire (2), 0.5 mm

2

Cu / AWG 20 with stranded drain wire (20) of shield

3 Insulated wire (3), 0.5 mm

2

Cu / AWG 20 with stranded drain wire (30) of shield

4 Outer sheath
5 Insulation layers
6 Stranded drain wire (6) for the outer shield (60), 0.5 mm

2

Cu / AWG 20 (not insulated, bare)