2. Basic Operation
9
2.2 How to Make the Reading More Readable
This function is useful when you want to obtain averaged readings as readings were volatile.
* This function is not available in Average Value Measurement Mode, Intermittent Operation Measurement Mode
and Air Flow Measurement Mode.
Mode
How to Take in Measurement Data
Explanation
FAST
Data is taken 4 times for one second. The
average value of the 4 data is displayed as
an instantaneous value every second.
SLOW 1
The average value for 5 seconds is
displayed every second.
Data shifts by 1 second.
SLOW 2
The average value for 10 seconds is
displayed every second.
Data shifts by 1 second.
0 5 10 15 20sec
(measuring time)
average for 1 sec
average for 5 sec
0 5 10 15 20sec
(measuring time)
average for 10 sec
0 5 10 15 20sec
(measuring time)
★
Every time you press FAST/SLOW key, the display at the
upper right corner changes in sequence.
Press FAST/SLOW key.
08.07.01 FAST
10:26:40
15.7 m/s
24.6 ℃
FAST:
Instantaneous value every second
SLOW1: Moving average deviation for 5 sec
SLOW2: Moving average deviation for 10 sec
FAST SLOW 1 SLOW 2