Probe – KANOMAX 6162 Anemomaster User Manual
Page 12
1. Getting Started
6
1.5 Getting Ready for Measurement
1.5.1 Connecting a Probe
Probe is calibrated together with the provided connection cable.
Please make sure the connection cable is connected when
performing a measurement.
* To change the length of a connection cable, recalibration
will be required.
1.5.2 Checking the Probe Number
Confirm that the number indicated on the one end of the probe
cable (Teflon cable) is consistent with the one indicated on the
screen of the instrument.
<
The probe number needs to be checked when multiple probes are purchased or the same probe is to be
used with multiple Anemomasters or a spare probe is used.
>>> Calibration data of the probe is written on characteristic ROM installed on the bottom of the
instrument. Please check the probe number because each probe has different characteristic.
1.5.3 To Attach and Detach the Probe Board
(2) Loosen up the screw.
>>> The screw will not come out.
(3) Pull it up vertically.
As the board is connected to the connector,
do not twist it or pull only one side.
>>> It may cause the connection failure.
(1) Open the bottom lid.
aaaaaaaaaa
Connection Cable
(heat-resistant: 80 °C)
Probe Cable
heat-resistant: 200°C
Probe Number
08.07.15
10:26:40
PROBE
46-H001
* This screen will be displayed when the
instrument is turned ON without the
probe connected to the instrument.
Probe Number
Are they consistent with each other?
Inside the bottom lid