Troubleshooting

§ 12

Troubleshooting

Problems

Possible Cause / Solution

Refer To

Display

does

not

appear when power is
turned ON.

Battery is inserted in wrong polarity.

Turn off the power and insert the battery

correctly.

Page 11
& 12

The battery is low.

Turn off the power and replace the

batteries.

Page 11
& 12

blinks

The battery is low.

Turn off the power and replace the

batteries.

Page 11
& 12

Reading is displayed
as ”-----”

Measurable range is exceeded.

The instrument must be used in the

specified measurement range.

Page 27

Probe wire disconnection or sensor damage.

Contact your local distributor for repair.

Page 34

Incorrect air velocity
reading.

Confirm that the direction mark of the probe
is facing against the airflow direction.

Page 14

Sensor of the probe is dirty.
Turn off the power, and clean the sensor.

Page 26

High air temperature
readings.

Correct reading cannot be obtained when
there is no airflow.

Minimum 0.1 m/s velocity is required for
measurement.

Page 14

Air velocity reading
does not change from
“0.0” when sampling in
Volumetric Flow / Air
Temperature
measurement mode.

Duct shape and dimension settings are not
made properly (setting is zero).

Provide correct duct shape/dimension

settings.

Page 14

Printing Failure

Confirm that the printer cable is connected
properly.

Page 22

Printer is not connected in the right order.

After connecting the printer, turn on the

instrument first, and then turn on the printer.

Page 22

Baud rate is not set properly.

Confirm the instrument and printer

settings.

Page 22

Data Transfer Failure

Confirm that the USB cable is connected
properly. Make sure that it is not confused
with the printer cable.

Page 25

Analog Output Failure

Confirm that the polarity of the output
terminal is correct.

Page 25

The reading is in “hold” mode.

Press the key to release the hold

mode.

Page 16

Incorrect Output Value

Load impedance is set lower than the
specified value.

Load impedance must be set to 5kΩ and

over.

Page 25