Troubleshooting, 12 troubleshooting – KANOMAX 6036 Multi-function thermal anemometer User Manual User Manual
Page 33

32
Troubleshooting
§ 12
Troubleshooting
Problems
Possible Cause / Solution
Refer To
Display
does
not
appear when power is
turned ON.
Battery is inserted in wrong polarity.
Turn off the power and insert the battery
correctly.
Page 11
& 12
The battery is low.
Turn off the power and replace the
batteries.
Page 11
& 12
blinks
The battery is low.
Turn off the power and replace the
batteries.
Page 11
& 12
Reading is displayed
as ”-----”
Measurable range is exceeded.
The instrument must be used in the
specified measurement range.
Page 27
Probe wire disconnection or sensor damage.
Contact your local distributor for repair.
Page 34
Incorrect air velocity
reading.
Confirm that the direction mark of the probe
is facing against the airflow direction.
Page 14
Sensor of the probe is dirty.
Turn off the power, and clean the sensor.
Page 26
High air temperature
readings.
Correct reading cannot be obtained when
there is no airflow.
Minimum 0.1 m/s velocity is required for
measurement.
Page 14
Air velocity reading
does not change from
“0.0” when sampling in
Volumetric Flow / Air
Temperature
measurement mode.
Duct shape and dimension settings are not
made properly (setting is zero).
Provide correct duct shape/dimension
settings.
Page 14
Printing Failure
Confirm that the printer cable is connected
properly.
Page 22
Printer is not connected in the right order.
After connecting the printer, turn on the
instrument first, and then turn on the printer.
Page 22
Baud rate is not set properly.
Confirm the instrument and printer
settings.
Page 22
Data Transfer Failure
Confirm that the USB cable is connected
properly. Make sure that it is not confused
with the printer cable.
Page 25
Analog Output Failure
Confirm that the polarity of the output
terminal is correct.
Page 25
The reading is in “hold” mode.
Press the key to release the hold
mode.
Page 16
Incorrect Output Value
Load impedance is set lower than the
specified value.
Load impedance must be set to 5kΩ and
over.
Page 25