Data storage and statistical calculation, Storing measured data, Data storage and statistical calculation procedure – KANOMAX 6036 Multi-function thermal anemometer User Manual User Manual
Page 19: 4 data storage and statistical calculation

18
Data Storage and Statistical Calculation
§ 4
Data Storage and Statistical Calculation
Storing Measured Data
Measured data can be stored in the built-in memory of the
instrument. The instrument can hold up to 1500 data records. When
storing the measured data, average, maximum and minimum
values will be calculated for the data group to be stored. Each data
group is stored with a storage number (shown as DNo xxx) which
starts from 001 for each mode. Contents of the data stored in each
measurement mode are shown in the following table.
Measurement Mode
Stored Data
Air Velocity /
Air Temperature
Air Velocity, Air Temperature and Data Storage No.
Volumetric Flow Rate /
Air Temperature
Volumetric Flow Rate, Air Temperature,
Data Storage No., and Duct Shape/Size
*Pressure /
Air Temperature
Pressure, Air Temperature, and Data Storage No.
You can print out the measured data by using the supplied software
“PC-LINK Software” or the optional printer.
Data Storage and Statistical Calculation Procedure
Data storage procedure is described below by taking an example of
an operation in the “Air Velocity / Air Temperature Measurement
Mode”. Data storage procedure is same in other mode as well.
●
Temporary Memory
By pressing the
key while
measuring, the current reading will be
stored in the temporary memory.
●
Number of Stored Data Records
Indicates the number of data records
stored in the temporary memory.
●
Temporary Data Storage
The displayed reading will be stored
in the temporary memory each time
the
key is pressed, and the
number of stored data records will
increase by one.
*
A maximum of 1500
data records can be stored.
Number stored data records
*
Temporary Memory – Data will be cleared once measurement
mode is changed, the power is turned off or calculation is executed.
*
[No. of data records stored] + [No. of data records that can
temporarily be stored] = 1500
■
Pressure mode is
available on
MODEL6035-B0/
6035-C0/
6036-B0/
6036-C0
■
Data storage
function is only
available on
MODEL6036