beautypg.com

HT instruments MPP300 User Manual

Page 21

background image

MPP300

EN - 19

22. At any time while recording it will be possible to

analyze its current status by selecting the button
Saved data management in the GENERAL
MENU
. The following information will be shown:

a. starting date and time of recording
b. the value set for the integration period
c. the number of periods elapsed from the

beginning of the recording

d. the remaining memory capacity for

recording

Press the ESC key to exit the screen

23. Now it is possible to bring the unit SOLAR-02 near the PV strings to measure irradiation and

temperature by means of the relevant probes. When the distance between unit SOLAR-
02 and MPP300 does not allow the RF connection, on the display of SOLAR-02 the
symbol “

” flashes for approx 30s and then disappears. Unit MPP300 steadily

searches for the RF connection with unit SOLAR-02.

24. Position the reference cell onto the surface of the PV modules. Please refer to the relevant

User Manual for a correct assembly.


25. Put the temperature sensor in contact with the rear side of the module and fasten it with

some tape; prevent touching it with your fingers (as this could alter the measure).


26. Wait for a few seconds to allow the probes to reach a steady measure and then connect the

irradiation probe to input PYRA/CELL and the temperature probe to input TEMP of unit
SOLAR-02.


27. Wait for the message “READY” to appear on the display of SOLAR-02 to indicate that the

unit has detected the data with solar irradiation > minimum threshold set (see SOLAR300N
user’s manual).


28. With the message “READY” shown on the display, wait for a time period at least equal to the

set integration period (see SOLAR300N user’s manual) in order to take a significant number
of samples (>=2).


29. Disconnect the irradiation and temperature probes from unit SOLAR-02 and bring the unit

near unit MPP300. The two units must be near each other (max distance 1m).

30. Connect (if disconnected) the SOLAR300N unit to the MPP300 unit. The MASTER LED

must always be flashing to indicate the connection between SOLAR300N and MPP300.


31. Press key

 on SOLAR-02 to activate the RF connection again. Consequently, the

REMOTE LED on unit MPP300 will start flashing.