Mpp300 – HT instruments MPP300 User Manual
Page 14
MPP300
EN - 12
17. The display of SOLAR I-V will show the values of the
general electrical parameters of the system being
measured
In particular, in this screen:
Pdc = General dc power (sum of the string powers)
Pac = ac power (if single-phase) or sum of the ac powers (if
three-phase)
We recommend checking that the values of the electrical
parameters (Pnom, Pdc, Pac) and of the ac performance
(
ac) are consistent with the system being measured (e.g.:
ac > 1 is not physically acceptable)
15/05/10 15:34:26
P R p
- - -
I r r
- - -
W / m 2
P n o m 3 . 5 0 0
k W
T c
- - -
° C
T e
- - -
° C
P d c
3 . 1 2 5
k W
P a c
2 . 9 6 0
k W
n d c
- - -
n a c
0 . 9 5
G O – s t a r t r e c
Select
M P P
18.
On SOLAR I-V, press key (
) to access the second
screen which contains the values of the output DC
parameters of the strings according to the number of DC
inputs set (see the User Manual of SOLAR I-V)
In particular, in this screen:
Vdcx = dc voltage of string x.
Idcx = dc current of string x.
Pdx = dc power of string x.
We recommend checking that the values of the electrical
parameters (Vdc, Idc, Pdc) are consistent with the system
being measured
15/05/10 15:34:26
V d c 1
4 6 0 . 1
k W
V d c 2 4 6 1 . 4
V
V d c 3
4 6 2 . 5
A
I d c 1
2 . 2 5
A
I d c 2
2 . 3 1
A
I d c 3
2 . 2 1
A
P d c 1
1 . 0 3 5
k W
P d c 2
1 . 0 6 6
k W
P d c 3
1 . 0 2 4
k W
G O – s t a r t r e c
Select
M P P
19. On SOLAR I-V, press key (
) to access the third screen
which contains the values of the electrical parameters on
the AC side of the inverter, consistently with the settings
made (see – SOLAR I-V user’s manual, single-phase,
three-phase 4 wires).
In particular, in this screen:
Vacxy = ac voltage between Phase and Neutral (if single-
phase) or between Phases x and y (if three-phase)
Iacx = ac current of phase x
Pacx = ac power of phase x
We recommend checking that the values of the electrical
parameters (Vac, Iac, Pac) are consistent with the system
being measured
15/05/10 15:34:26
V a c 1 2 4 0 1 . 4
V
V a c 2 3 4 0 1 . 1
V
V a c 3 1 4 0 0 . 1
V
I a c 1
4 . 2 6
A
I a c 2
4 . 2 6
A
I a c 3
4 . 2 7
A
P a c 1
9 8 7
W
P a c 2
9 8 6
W
P a c 3
9 8 5
W
G O – s t a r t r e c
Select
M P P
E x a m p l e o f a s c r e e n f o r
P V s y s t e m s w i t h t h r e e -
p h a s e o u t p u t
20. Keeping the three instruments near each other (max distance
approx. 1m), press the GO/STOP key on SOLAR I-V to start
testing. Consequently:
The display of SOLAR I-V shows the message “Rec. start
waiting”
The display of SOLAR-02 shows the message “HOLD” and
the time, expressed in seconds, remaining before the
recording is started
On MPP300, the STATUS LED turns on green (not flashing)
15/05/10 15:34:26
P R p
- - -
I r r
- - -
W / m 2
P n o m 3 . 5 0 0
k W
T c
- - -
° C
T e
- - -
° C
P d c
3 . 1 2 5
k W
P a c
2 . 9 6 0
k W
n d c
- - -
n a c
0 . 9 5
R e c . S t a r t W a i t i n g
Select
M P P