beautypg.com

GW Instek GDS-1000-U Series Quick start guide User Manual

Page 2

background image

8. Press the Autoset key. A

square waveform will

appear in the center of the

display.

9. Press the Display key, then

Type and select the vector

waveform type.

10. Turn the adjustment point on the probe to flatten

the square waveform edge.

Over

Compensation

Normal

Under

Compensation

11. Setting up the oscilloscope is complete. You may

start to use the oscilloscope.

S

PECIFICATIONS

The specifications apply when the oscilloscope is

powered on for at least 30 minutes under +20°C~+30°C.

Model Specific Specifications

GDS-1052-U
Bandwidth (–3dB)

DC coupling: DC ~ 50MHz

AC coupling: 10Hz ~ 50MHz

Bandwidth Limit

20MHz (−3dB)

Trigger Sensitivity

0.5div or 5mV (DC ~ 25MHz)

1.5div or 15mV

(25MHz~50MHz)

External Trigger
Sensitivity

~ 50mV (DC~25MHz)

~ 100mV (25MHz~50MHz)

Rise Time

< 14ns approx.

GDS-1072-U
Bandwidth (–3dB)

DC coupling: DC ~ 70MHz

AC coupling: 10Hz ~ 70MHz

Bandwidth Limit

20MHz (−3dB)

Trigger Sensitivity

0.5div or 5mV (DC ~ 25MHz)

1.5div or 15mV

(25MHz~70MHz)

External Trigger
Sensitivity

~ 50mV (DC~25MHz)

~ 100mV (25MHz~70MHz)

Rise Time

< 5.8ns approx.

GDS-1102-U
Bandwidth (–3dB)

DC coupling: DC ~ 100MHz

AC coupling: 10Hz ~

100MHz

Bandwidth Limit

20MHz (−3dB)

Trigger Sensitivity

0.5div or 5mV (DC ~ 25MHz)

1.5div or 15mV

(25MHz~100MHz)

External Trigger
Sensitivity

~ 50mV (DC~25MHz)

~ 100mV (25MHz~100MHz)

Rise Time

< 3.5ns approx.

Common Specifications

Vertical
Sensitivity

2mV/div~10V/Div (1-2-5

increments)

Accuracy

± (3% x |Readout|+0.1div +

1mV)

Bandwidth

See model-specific

specifications

Rise Time

See model-specific

specifications

Input Coupling

AC, DC, Ground

Input Impedance

1MΩ±2%, ~15pF

Polarity

Normal, Invert

Maximum Input

300V (DC+AC peak), CAT II

Math Operation

+, –, FFT

Offset Range

2mV/div~50mV/div: ±0.4V

100mV/div~500mV/div:

±4V

1V/div~5V/div: ±40V

10V/div : ±300V

Trigger
Sources

CH1, CH2, Line, EXT

Modes

Auto, Normal, Single, TV,

Edge, Pulse

Coupling

AC, DC, LF rej, HF rej, Noise

rej

Sensitivity

See model-specific

specifications

External Trigger
Range

DC: ±15V, AC: ±2V

Sensitivity

See model-specific

specifications

Input Impedance

1MΩ±2%, ~15pF

Maximum Input

300V (DC+AC peak), CATII

Horizontal
Range

1ns/div~50s/div, 1-2.5-5

increment

Roll: 50ms/div – 50s/div

Modes

Main, Window, Window

Zoom, Roll, X-Y

Accuracy

±0.01%

Pre-Trigger

10 div maximum

Post-Trigger

1000 div

X-Y Mode
X-Axis Input

Channel 1

Y-Axis Input

Channel 2

Phase Shift

±3° at 100kHz

Signal Acquisition
Real-Time

250M Sa/s maximum

Equivalent

25G Sa/s maximum

Vertical Resolution

8 bits

Record Length

4k points Maximum

Acquisition

Normal, Peak Detect,

Average

Peak Detection

10ns (500ns/div ~ 50s/div)

Average

2, 4, 8, 16, 32, 64, 128, 256

Cursors and Measurement
Voltage

Vpp, Vamp, Vavg, Vrms,

Vhi, Vlo, Vmax, Vmin, Rise

Preshoot/ Overshoot, Fall

Preshoot/ Overshoot

Time

Freq, Period, Rise Time, Fall

Time, + Width, – Width,

Duty Cycle

Cursors

Voltage difference (ΔV) and

Time difference (ΔT) between

cursors

Auto Counter

Resolution: 6 digits,

Accuracy: ±2%

Signal source: All available

trigger source except the

Video trigger

Control Panel Function
Autoset

Automatically adjust Vertical

Volt/div, Horizontal

Time/div, and Trigger level

Save/Recall

Up to 15 sets of measurement

conditions and waveforms

Display
LCD

5.7 inch, TFT, brightness

adjustable

Resolution (dots)

234 (Vertical) x 320

(Horizontal)

Graticule

8 x 10 divisions

Display Contrast

Adjustable

Interface
USB Slave Connector

USB1.1 & 2.0 full speed

compatible (flash disk not

supported)

USB Host connector

Image (BMP) and waveform

data (CSV)

Probe Compensation Signal
Frequency range

1kHz ~ 100kHz adjustable,

1kHz step

Duty cycle

5% ~ 95% adjustable, 5% step

Amplitude

2Vpp±3%

Power Source
Line Voltage

100V~240V AC, 47Hz~63Hz

Power Consumption

18W, 40VA maximum

Fuse Rating

1A slow, 250V

Operation Environment
Storage Temperature

-20°C~70°C, no condensation

Relative humidity

80% @ 70°C

Dimensions and Weight
Dimensions

310(W) x 142(H) x 140(D) mm

Weight

Approx. 2.5kg

Probe Specifications

GTP-070A-4 (GDS-1052-U, GDS-1072-U)
Probe Position

Position x10 Position x1

Attenuation Ratio

10:1

1:1

Bandwidth

DC ~ 70MHz DC~6MHz

Input Resistance

10MΩ when

used with

1MΩ input

1MΩ when

used with

1MΩ input

Input Capacitance

28pF~32pF

120pF ~220pF

Maximum Input Voltage

≤600Vpk,

Derating with

frequency

≤200Vpk,

Derating with

frequency

Temperature

–10°C ~ 50°C

Relative Humidity

≤85% @35°C

Safety Standard

EN 61010-031 CAT II

GTP-100A-4 (GDS-1102-U)
Probe Position

Position x10 Position x1

Attenuation Ratio

10:1

1:1

Bandwidth

DC ~ 100MHz DC~6MHz

Input Resistance

10MΩ when

used with

1MΩ input

1MΩ when

used with

1MΩ input

Input Capacitance

14.5~17.5pF

approx.

85~115pF

approx.

Maximum Input Voltage

≤600Vpk,

Derating with

frequency

≤200Vpk,

Derating with

frequency

Temperature

–10°C ~ 50°C

Relative Humidity

≤85% @35°C

Safety Standard

EN 61010-031 CAT II

Dimensions

VI

D/

E

MI

T

VI

D/

ST

L

O

V

VI

D/

ST

L

O

V

U

N

E

M

U

N

E

M

2

H

C

HT

A

M

1

H

C

A

cq

uir

e

Di

sp

la

y

U

tili

ty

H

elp

R

un

/S

to

p

VA
R

IAB
LE

FO

RC
E

A

uto

se

t

C

urs

or

S

IN

G

LE

H

ard

cop

y

M

ea

su

re

S

ave

/R

eca

ll

LEVE
L

RE

G

GI

RT

L

AT

N

OZ

I

R

O

H

LA

CI

T

RE

V

CH
1

C

AT

30

0V M

15

pF

M

AX

. 3

00

Vp

k

1

CH
2

EX

T

TR

IG

C

AT

30

0V M

15

pF

M

AX

. 3

00

V

pk

1

Y

X

310.

0

341.

5

142.0

162.3

115.

1

14

9.

0

159

.0

150
MH
z

1

G

Sa

/s

D

igi

ta

l St

orage
O

sc

illo

sc

op

e

G

D

S-1

15

2A

-U

EC Declaration of Conformity

We

GOOD WILL INSTRUMENT CO., LTD.

No.7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236,

Taiwan

GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.

No. 69, Lushan Road, Suzhou New District Jiangsu, China

declares that the below mentioned product

GDS-1052-U, GDS-1072-U, GDS-1102-U

Are herewith confirmed to comply with the requirements set

out in the Council Directive on the Approximation of the Law

of Member States relating to Electromagnetic Compatibility

(2004/108/EC) and Low Voltage Equipment Directive

(2006/95/EC). For the evaluation regarding the

Electromagnetic Compatibility and Low Voltage Equipment

Directive, the following standards were applied:

◎ EMC

EN 61326-1 :
EN 61326-2-1:

Electrical equipment for measurement, control and
laboratory use –– EMC requirements (2006)

Conducted and Radiated
Emissions
EN 55011: 2009+A1:2010

Electrostatic Discharge
EN 61000-4-2: 2009

Current Harmonic
EN 61000-3-2:
2006+A1:2009+A2:2009

Radiated Immunity
EN 61000-4-3:
2006+A1:2008+A2:2010

Voltage Fluctuation
EN 61000-3-2: 2008

Electrical Fast Transients
EN 61000-4-4: 2004+A1:2010

-------------------------

Surge Immunity
EN 61000-4-5: 2006

-------------------------

Conducted Susceptibility
EN 61000-4-6: 2009

-------------------------

Power Frequency Magnetic Field
EN 61000-4-8: 2010

-------------------------

Voltage Dips/ InterruptsIEC
EN 61000-4-11: 2004

◎ Safety

Low Voltage Equipment Directive 2006/95/EC
Safety Requirements
IEC/EN 61010-1: 2001