beautypg.com

Modular mechanical design, 2 modular mechanical design, The dm-100 o – Detcon DM-100 User Manual

Page 6: Figure 3 construction of galvanic cell, Figure 4 sensor assembly breakaway

background image

Model DM-100

DM-100 Instruction Manual

Rev. 3.3

Page 2 of 40

The DM-100 O

2

deficiency sensor technology is a two electrode galvanic metal air battery type cell, which is

housed as a field replaceable intelligent plug-in sensor. The cell is diffusion limited and functions as a direct
current generator proportional to the amount of oxygen adsorption. The sensors are temperature compensated
and show good accuracy and stability over the operating temperature range of -20° to 50°C (-4° to +122° F).
The sensor is warranted for two years and has an expected service life of up to 2.5 years in ambient air at
20.9% oxygen.

Figure 3 Construction of Galvanic Cell

1.2 Modular Mechanical Design

The Model DM-100 Sensor Assembly is completely modular and is made up of four parts (Figure 4):

1. DM-100 Intelligent Transmitter Module (ITM)
2. Intelligent Plug-in Sensor (varies by gas type and range)
3. 100 Series Bottom Housing Assembly (H2S assemblies use 100 Series Bottom Housing Assembly

with Integral Filter)

4. Splash Guard.

O-Ring

100 Series Bottom

Housing Assembly

Plug-in
Replaceable
Sensor

Interconnect

Wires

Housing Bottom

Locking Screw

Intelligent Transmitter Module (ITM)
Micro-processor controlled circuit

encapsulated in an Explosion proof
housing.

Magnetic

Programming

Switches

Splash Guard

D

E

T

C

O

N

,

IN

C

.

H

2

S

1

0

0

P

P

M

D

M

-1

0

0

3

4

" NPT Locking Nut

Figure 4 Sensor Assembly Breakaway

NOTE

All metal components are constructed from electro polished 316 Stainless Steel in order to
maximize corrosion resistance in harsh environments.