5. Troubleshooting Guide

If the ITM detects any functional errors the ITM will bring the mA output to 0mA and hold it there until the
problem has been resolved. If the unit has a Display, the display will show an error code. Error codes are
displayed as FLXX where XX is the actual error code.

The Display Error Codes are:
FL01

Auto Span Fault

FL05

ITM Memory Fault

FL06

Processor Fault

FL07

Clearing Fault

FL08

Stability Fault

FL09

Range Fault

FL10

Sensor Fault

FL11

Zero Fault

FL12

Sensor Fault 2

AutoSpan Fault (FL01)

This fault occurs when a span calibration has not been performed in the last 180 days. Execute a span
calibration to clear this fault.

Transmitter Not Responding (FL05, FL06)

 Verify condulet has no accumulated water or abnormal corrosion.

 Verify required DC power is applied to correct terminals. Swap with a known-good ITM to determine

if ITM is faulty.

Clearing Problem (FL07)

 The sensor must recover to < 10% of range in < 5 min after Span calibration is complete. Verify the

cal gas is being removed at the appropriate time during span calibration.

 Background gas may be present. Use bottled air (zero air or N

2

) if there is a known continuous

background level.

 Check validity of span gas using pull tube or other means (check MFG date on cal gas cylinder).

Use proper cal gas regulators and tubing for highly corrosive gases.

 Check for obstructions affecting cal gas hitting sensor face, including being wet, blocked, or corroded.

H2S sensors assemblies use 100 Series Bottom Housing Assembly with integral filter. Clean or replace

if necessary.

 Replace the plug-in toxic sensor.

Stability Fault (FL08)

 Verify the cal gas cylinder is not close to being empty.

 Check validity of span gas using pull tube or other means, verify MFG date on cal gas cylinder. Use

proper cal gas regulators and tubing for highly corrosive gases.

 If using Splashguard with Integral Cal Port, must use Calibration Wind Guard or air movement can

compromise span gas delivery

 Check for obstructions affecting cal gas hitting sensor face, including being wet, blocked, or corroded.

H2S sensors assemblies use 100 Series Bottom Housing Assembly with integral filter. Clean or replace

if necessary.

 Replace the plug-in toxic sensor.