Bit 6

1-Alarm 1 is ascending

0-Alarm 1 is descending

Bit 5

1-Alarm 1 is normally energized

0-Alarm 1 is normally de-energized

Bit 4

1-Alarm 1 is latching

0-Alarm 1 is non-latching

Bit 3

1-Alarm 1 Relay is energized

0-Alarm 1 Relay is not energized

Bit 2

1-Fault is normally energized

0-Fault is normally de-energized

Bit 1

1-Fault is latching

0-Fault is non-latching

Bit 0

1-Fault Relay is energized

0-Fault Relay is not energized

The following is a typical Master Query for device # 8:

Field Name

HEX

DEC

RTU

Slave Address

08

8

0000 1000

Function

03

3

0000 0011

Start Address Hi

00

0

0000 0000

Start Address Lo

00

0

0000 0000

No. of Registers Hi

00

0

0000 0000

No. of Registers Lo

06

6

0000 0110

CRC

##

#### ####

CRC

##

#### ####

The following is a typical Slave Response from device # 8:

Field Name

HEX

DEC

RTU

Slave Address

08

8

0000 1000

Function

03

3

0000 0011

Byte Count

0C

12

0000 1100

Reg40000 Data Hi

02

2

0000 0010

Reg40000 Data Lo

64

100

0110 0100

Reg40001 Data Hi

00

0

0000 0000

Reg40001 Data Lo

64

100

0110 0100

Reg40002 Data Hi

00

0

0000 0000

Reg40002 Data Lo

07

7

0000 0111

Reg40003 Data Hi

00

0

0000 0000

Reg40003 Data Lo

0A

10

0000 1010

Reg40004 Data Hi

00

0

0000 0000

Reg40004 Data Lo

14

20

0001 0100

Reg40005 Data Hi

05

5

0000 0101

Reg40005 Data Lo

50

80

0101 0000

CRC

##

#### ####

CRC

##

#### ####

Additional Notes:

The calibration LED will light when the transmitter is sending a response to a Master Query.

Communications are 9600 baud, 8 data bits, 1 stop bit, No parity, half duplex 485.

3.12 D

ISPLAY

C

ONTRAST

A

DJUST

Model IR-640/IR-642 MicroSafe™ sensors feature a 16 character backlit liquid crystal display. Like most LCDs,

character contrast can be affected by viewing angle and temperature. Temperature compensation circuitry included

in the MicroSafe™ design will compensate for this characteristic, however temperature extremes may still cause a

shift in the contrast. Display contrast can be adjusted by the user if necessary. However, changing the contrast

requires that the sensor housing be opened, thus declassif ication of the area is required.

To adjust the display contrast, remove the enclosure cover and use a jewelers screwdriver to turn the contrast

Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.20