beautypg.com

Detcon IR-642 User Manual

Page 18

background image

If during calibration the sensor circuitry is unable to attain the proper adjustment for zero or span, the sensor will

enter into the calibration fault mode and cause the display to alternate between the sensor’s current status reading and

the calibration fault screen which appears as:

“CAL FAULT.2”.

The following conditions will cause a calibration fault:

1 - Zero calibration cannot converge.

2 - Auto span cannot converge (too noisy or too unstable).

3 - Span gas is not applied before 1 minute elapses.

Fail-Safe/Fault Supervision

Model IR-640/IR-642 MicroSafe™ sensors are programmed for fail-safe operation. All of the fault conditions listed

below will activate the fault relay, illuminate the fault LED, cause the mA output to drop to zero (0) mA, and cause

the display to read its corresponding fault condition.

Memory Error

If the processor can’t save values to memory, the display will indicate:

“MEMORY ERROR”.

Zero Fault

If the sensor should drift below –0.02%, the display will indicate:

“ZERO FAULT”.

Lamp Fault

If the lamp signal is lost, the display will indicate:

“LAMP FAULT.2”.

Reference Peak High Fault

If the reference peak signal is too high (>3600), the display will indicate:

“SIGNAL FAULT.31”.

Active Peak High Fault

If the active peak signal is too high (>3600), the display will indicate:

“SIGNAL FAULT.32”.

Reference Peak Low Fault

If the reference peak signal is too low (<500), the display will indicate:

“SIGNAL FAULT.41”.

Active Peak Low Fault

If the active peak signal is too low (<500), the display will indicate:

“SIGNAL FAULT.42”.

Reference Peak to Peak Low Fault

If the reference peak to peak signal is too low (<200), the display will indicate:

“SIGNAL FAULT.51”.

Active Peak to Peak Low Fault

If the active peak to peak signal is too low (<200), the display will indicate:

“SIGNAL FAULT.52”.

3.11 RS-485 P

ROTOCOL

Detcon MicroSafe™ sensors feature Modbus™ compatible communications protocol and are addressable via rotary

dip switches for multi-point communications. Other protocols are available. Contact the Detcon factory for specific

protocol requirements. Communication is two wire, half duplex 485, 9600 baud, 8 data bits, 1 stop bit, no parity,

with the sensor set up as a slave device. A master controller up to 4000 feet away can theoretically poll up to 256

different sensors. This number may not be realistic in harsh environments where noise and/or wiring conditions

would make it impractical to place so many devices on the same pair of wires. If a multi-point system is being uti-

lized, each sensor should be set for a different address. Typical address settings are: 01, 02, 03, 04, 05, 06, 07, 08, 09,

0A, 0B, 0C, 0D, 0E, 0F, 10, 11, etc.

In most instances, RS-485 ID numbers are factory set or set during installation before commissioning. If required,

the RS-485 ID number can be set via rotary dip switches located on the preamp circuit board. However, any change

Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.18

This manual is related to the following products: