7

Reset

Reset (soft reset) restores factory settings and returns the Gage to

a known condition. The following occurs:
- All batches, stored measurements, images, batch names and

screen captures are erased.

- All calibration and range adjustments are cleared and returned

to the Gage’s factory settings.

- Menu settings are returned to the following:

Memory = OFF

Graphics = ON

Cal Lock = OFF

Bluetooth = OFF

Display = None

Perform a more thorough Hard Reset by powering down the

Gage, waiting several seconds, then simultaneously holding both

the center and (+) buttons until the Reset symbol

appears. This returns the instrument to a known, “out-of-the-box”

condition. It performs the same function as a menu Reset with the

addition of:

Date, Time and WiFi are not affected by either Reset.

Units = microns

Flip Display = Normal

White on Black = OFF

Language = English

Battery Type = Alkaline

Backlight = Normal

USB Drive= ON

Auto Sync = OFF

- Bluetooth Pairing info is cleared.

- Menu settings are returned to the following:

Setup Menu

Setup Menu

Selects the type of batteries used in the Gage from a choice of

“Alkaline”, “Lithium” or “NiMH” (Nickel-metal hydride

rechargeable). If NiMH is selected, the Gage will trickle charge the

batteries while connected via USB to a PC or optional AC charger.

The battery state indicator icon is calibrated for the selected

battery type. No damage will occur if the wrong battery type is

selected.

DeFelsko recommends the use of eneloop (NiMH)

rechargeable batteries.

Battery Type

Layers = 1

Layer Name = Layer 1

Layer Color = blue

NOTE:

NOTE: