Cal settings menu – DeFelsko PosiTector 200 v.3.0 Quick Guide User Manual

Page 6

background image

5

The PosiTector 200 interprets the largest “echo” or “echoes”

within the selected range as the coating/coating or

coating/substrate echo. If the coating thickness is outside this

range, incorrect or dashed readings may occur.

IMPORTANT:

Cal Settings Menu

Cal Settings Menu

The probe must be periodically zeroed using the Zero menu option

to compensate for both extreme temperature and probe wear

effects. Before using, allow the probe to reach ambient

temperature. Wipe the probe clean of couplant. The icon

appears after probe is zeroed.

If measurements will be made in extreme hot or cold temperatures,

it is recommended to Zero the probe in the working environment.

If measurements will be made on rough substrates, it is

recommended to periodically Zero the probe to compensate for

wear.

Zero

Set Range

The measuring range of each probe can be changed depending on

the specific application or the expected thickness range of the

coating system.

For most applications, the default range values do not have to be

adjusted. But some conditions, like surface roughness, may cause

the Gage to display very low or non-repeatable readings. In this

case, the low range may be increased to cause the Gage to only

display readings above the Lo value set by the user.

To adjust Set Range values...

1. Select

or .

2. Use the (-)(+) buttons to decrease/increase the displayed value.