Setup menu – DeFelsko PosiTector 6000 v.7.3 Quick Guide User Manual
Page 5

4
Reset
Reset (soft reset) restores factory settings and returns the
instrument to a known condition. The following occurs:
- All batches, stored measurements, images, and batch names
are erased.
- All calibration adjustments and Cal Memory are cleared and
returned to the Gage’s factory calibration settings.
- Menu settings are returned to the following:
Memory = OFF
High Res = OFF
Statistics Mode = OFF
Hi Lo Alarm = OFF
Scan Mode = OFF
Cal Lock = OFF
Bluetooth = OFF
Display = None
N Lock = OFF
Setup Menu
Setup Menu
2 Pt Adjust
Provides greater accuracy within a limited, defined range. Measure
a thin value (often zero) and a thicker value. These values should
bracket your expected thickness range.
When checked, the icon appears and the current Cal settings
are “locked” to prevent further user adjustments.
Use when operating regularly on non-ferrous substrates. The
icon appears and the probe will only use the eddy current principle
to shorten measurement time and extend battery life. Useful when
measuring coatings over plated steel.
The current cal setting (ex. Cal 3) displays in the upper right corner
of the LCD. A stored cal setting can be restored when you return
to a particular part. Cal 1 can be adjusted but never deleted, and
is always made active with factory settings after a Reset (pg. 4). A
new cal setting cannot be created if a batch is open with readings.
(FN ferrous/ non-ferrous cobmination probes only)
Cal Memory
(Advanced models only)
Cal Lock
N Lock