Setup menu – DeFelsko PosiTector 6000 v.7.3 Quick Guide User Manual

Page 5

background image

4

Reset

Reset (soft reset) restores factory settings and returns the

instrument to a known condition. The following occurs:

- All batches, stored measurements, images, and batch names

are erased.

- All calibration adjustments and Cal Memory are cleared and

returned to the Gage’s factory calibration settings.

- Menu settings are returned to the following:

Memory = OFF

High Res = OFF

Statistics Mode = OFF

Hi Lo Alarm = OFF

Scan Mode = OFF

Cal Lock = OFF

Bluetooth = OFF

Display = None

N Lock = OFF

Setup Menu

Setup Menu

2 Pt Adjust

Provides greater accuracy within a limited, defined range. Measure

a thin value (often zero) and a thicker value. These values should

bracket your expected thickness range.

When checked, the icon appears and the current Cal settings

are “locked” to prevent further user adjustments.

Use when operating regularly on non-ferrous substrates. The

icon appears and the probe will only use the eddy current principle

to shorten measurement time and extend battery life. Useful when

measuring coatings over plated steel.

The current cal setting (ex. Cal 3) displays in the upper right corner

of the LCD. A stored cal setting can be restored when you return

to a particular part. Cal 1 can be adjusted but never deleted, and

is always made active with factory settings after a Reset (pg. 4). A

new cal setting cannot be created if a batch is open with readings.

(FN ferrous/ non-ferrous cobmination probes only)

Cal Memory

(Advanced models only)

Cal Lock

N Lock