beautypg.com

Cal settings menu – DeFelsko PosiTector 6000 v.7.3 Quick Guide User Manual

Page 4

background image

3

Probes are factory calibrated and perform an automatic self-check

each time a measurement is taken. For many applications no

further adjustment is necessary after a Reset (pg. 4). Just check

ZERO on the uncoated substrate, then measure. However,

sometimes Gage readings can be influenced by changes in

substrate shape, composition, surface roughness or by measuring

in a different location on the part. That is why Cal adjustments are

made possible. The XX symbol disappears whenever a Cal

adjustment is made to the Gage.
Where a Cal adjustment method has not been specified, use a

1-point method first. If measuring the included shims on your

surface reveals inaccuracies, use the 2-point method. Factory Cal

settings can be restored at any time by performing a Reset (pg. 4),

creating a NEW Cal setting (See Cal Memory pg. 4), or by

DELETING the adjustments made to the Cal 1 calibration setting

(pg. 4). The XX symbol appears on the display whenever factory

Cal settings are in use.
With “FN” Gages, calibration adjustments are made only to the “F”

or “N” mode (stored independently under a particular Cal),

whichever was measured last.

Cal Settings Menu

Cal Settings Menu

1-Pt Cal Adjustment- Also known as an offset

or correction value, there are 4 ways to perform this:

(i) Simple Zero Adjustment- Measure your uncoated part. Lift

the probe and adjust to "0" using the (+) and (-) buttons.
(ii) Average Zero Adjustment

A preferred method

to (i) is to average 3 to 10 measurements on the uncoated part.
(iii) Simple Adjustment to a Known Thickness- Measure a

shim of known thickness placed on the uncoated part. Lift the

probe and adjust to the shim thickness using the (+) and (-)

buttons.
(iv) Average Adjustment to a Known Thickness

A preferred method to (iii) is to average 3 to 10 measurements

of a shim of known thickness placed on the uncoated part using

the (+) and (-) buttons.

1 Pt Adjust

Zero

1 Pt Adjust