To disconnect a probe from a body, power-down the

instrument and slide the plastic probe connector

horizontally (in the direction of the arrow) away from the

body. Reverse these steps to attach a new probe.
When powered-up the PosiTector automatically

determines what type of probe is attached and does a

self-check.

Adjust Down

Access the Menu

Adjust Up

Restore brightness

after dimming

(Advanced only)

Button Functions- Normal Operation

Probes

Probes

Additionally, the PosiTector accepts a number of probe types

including magnetic and eddy current coating thickness, surface

profile, environmental and ultrasonic wall thickness probes.

For the latest information on probe interchangeability,

see www.defelsko.com/probes

Calibration, Verification, and Adjustment

Calibration, Verification, and Adjustment

The PosiTector 6000 non-destructively measures the thickness of

coatings on metals.Three steps ensure best accuracy…

1. Calibration: typically done by the manufacturer or qualified lab.

All probes include a Certificate of Calibration.

2. Verification of Accuracy: as done by the user on known

reference standards such as the included plastic shims or

optional coated thickness standards.

3. Adjustment: Adjustment, or Cal Adjustment is the act of

aligning the Gage’s thickness readings to match that of a known

sample in order to improve the effectiveness of the Gage on a

specific surface or in a specific portion of its measurement

range. 1-point or 2-point Cal adjustments are possible.