COSA Xentaur COSA 9610 User Manual
Page 36
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™
Page 36 of 44
NOTE
: Set oven temp to 81 2ºC & sample system to 50ºC for standard units,
90ºC for mid-temp models
3.4.2.
Adjustment procedure temperature regulator
This procedure describes the setting of the temperature controller. The controller
measures the temperature with the help of a K-type thermocouple. The heating is
arranged by a pulsing 5VDC signal. The on/off relation decides the added capacity. The
tuning program in the regulator will optimize the regulator algorithm. This program
decides the most optimal settings for the PID regulation.
Procedure:
1: Check the connections according to drawing.
2: Turn on the voltage.