beautypg.com

COSA Xentaur COSA 9610 User Manual

Page 18

background image

INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™

Page 18 of 44

3. IN OPERATION

3.1.

S

TART

-U

P

S

AMPLE

C

ONDITIONING

S

YSTEM

The sample conditioning system is located in the left-hand compartment. This chapter

describes how the components of the sample conditioning system should be set up, in

correct order, so a perfect start-up of the total analyzer system can be achieved. The

instructions hereunder should be performed step by step.

3.1.1.

Inspection, visual and external connections

Perform a visual inspection of the system and close all shut-off valves in the system.

Check the connecting fittings of the supply tubes to be correctly fitted and are not

leaking. This can be checked quite simple by unscrewing the nut from the fitting and

then check if "Front and Back ferrule" of the fitting are able to rotate but cannot be

moved in an axial direction. If this is not the case, this tube must be renewed. Then turn

the nut by hand and afterwards tighten it a 1/4 turn with a suitable spanner. The supply

and drainage tube can now be connected.

Because the supply line is under pressure and has been closed off on the COSA 9610™

side, the connecting fittings can be squirted with soap in order to detect any possible leaks.

When bubbles appear this indicates a leak and the tube concerned must immediately be

closed off at the supply point. Inspect the fitting and tube, replace components when

necessary.

3.1.2.

Air orifice selection

Before the COSA 9610™ is set into operation, the range of measurement must be

established. The air orifice is selected on the basis of the desired range of measurement.

The COSA 9610™ is standard equipped with the following orifices:


Gas: .0079 in to .0118 in (0.2 mm and 0.3mm)

Air: 0.55/0.60/0.65/0.70/0.75 for a dual range version extended with 0.80/0.85 and

0.90.