2.3.2.
COSA 9610™ in general purpose execution (Type 01 & 02)
The COSA 9610™ is to be mounted against an even wall or structural steel construction.
Fixing lugs are located on each corner of the cabinet. Fixings used must be suitable for
the weight of the COSA 9610™ (±331 lbs/150kg).
The COSA 9610™ must be mounted on such a level above the floor or underneath
located obstacles that the oven drain can be connected to a drain header or a condense
bottle.
The COSA 9610™ can optionally be supplied on a 304SS freestanding frame.
This frame is to be placed on a flat surface (i.e. concrete slab). Two holes
in the base of the frame enable the COSA 9610™ to be fixed to the floor .
2.3.3.
COSA 9610™ in explosion proof execution (Type 01-Ex & 02-Ex)
The COSA 9610™ is supplied on a 304SS freestanding frame. This frame
is to be placed on a flat surface (i.e. concrete slab). Two holes in the base of the
frame enable the COSA 9610™ to be fixed to the floor.