3 sterilizing, Function check – Seca 665 User Manual
Page 96

96 •
3. Please take note of the following terms:
8.3
Sterilizing
The sterilisation of this device is not permitted.
9. FUNCTION CHECK
Perform a function check before each use.
A complete function check includes:
• a visual inspection for mechanical damage
• a visual inspection and function check of the display
• function check of all the controls shown in the
section entitled "Overview" on page 65
• a function check of the optional accessories
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from page 97.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 97, you may not use the device.
– Have the device repaired by seca service or
by an authorized service partner.
– Please see the section entitled
"Servicing/recalibration" on page 99.
Interval
Component
Prior to every measurement
with direct skin contact
Weighing platform
After every measurement
with direct skin contact
Weighing platform
If required
• Display
• Membrane keyboard