Operation • 77

English

5. Set up the column:

a) undo the locking screw
b) keep the locking screw pulled out

c) swivel the column
d) allow the locking screw to engage
e) tighten the locking screw fully

6. Restore the power supply.

Aligning the scale

ATTENTION!

Incorrect measurement due to force shunts

If the scale and housing is in contact with some-
thing, e.g. a towel, weight will not be measured
correctly.
– Set the scale up so that only its foot screws

are in contact with the floor.

1. Place the scale on firm, level surface.

2. Undo the knurled wheels.

3. Level the device by turning the foot screws.

The air bubble in the spirit level must be located in
the exact center of the circle.

4. Tighten the knurled wheels in the direction of the

arrow.
The foot screws are secured against being
adjusted.

a)

b)

c)

d)

e)

2.

3.

4.

Spirit level

Air bubble