98 •

Only the "SYS" item is
visible in the "rF" menu.

• The wireless module is deactivated.

- Activate wireless module (see "Activating the

wireless module (SYS)" on page 91).

Only the "SYS" and "Lrn"
items are visible in the
"rF" menu.

• The wireless module is activated and no

wireless group is set up.
- Set up wireless group (see "Setting up a

wireless group (Lrn)" on page 91).

The items "APrt" and
"Time" are not visible in
the "rF" menu.

• No wireless printer registered to the wireless

group.
- Register the wireless printer to the wireless

group via the "Lrn" menu item (see "Setting
up a wireless group (Lrn)" on page 91).

After calling up the
menu, the "rF" item is
not displayed.

• The scale’s wireless module is defective.

- Inform seca service.

The display "Er:X:11"
appears.

The scale has too high a load or too high a load in
one corner.

- Take the load off the scale or distribute the

weight more evenly.

- Restart the scale.

The display "Er:X:12"
appears.

The scale has been switched on with too high a
load.

- Remove the load from the scale.
- Restart the scale.

The display "Er:X:16"
appears.

The scale was caused to oscillate, the zero point
could not be determined.

- Restart the scale.

The Enter key
(send/print) is pressed
and the display "Er:X:71"
appears.

No data transmission possible, wireless module
deactivated.

- Activate wireless module (see "Activating the

wireless module (SYS)" on page 91).

The Enter key
(send/print) is pressed
and the display
"Er:X:72" appears.

No data transmission possible, no wireless group
set up.

- Set up wireless group (see "Setting up a

wireless group (Lrn)" on page 91).

Problem

Cause / Solution