11.2 Weighing data
Power consumption
• with wireless module disabled
• with wireless module activated
approx. 32 mA
approx. 50 mA
Maximum runtime in battery mode
• with wireless module disabled
• with wireless module activated
approx. 5.600 minutes
Power supply unit recom-
mended
Medical device according to directives 93/42/
EEC and 2007/47/EC
Class I with measuring function
Calibration according to directive 2009/23/EC Class III
EN 60 601-1:
• Class II totally insulated appliance:
• Type B electromedical device:
Radio transmission
• Frequency band
• Transmission power
• Observed standards
2,433 GHz - 2,480 GHz
< 10 mW
EN 30028, EN 301 489-1, -17
Technical data for
seca 704
Weighing data
Calibration according to directive 2009/23/EC Class III
Maximum load
• Weighing range 1
• Weighing range 2
150 kg
300 kg
Minimum load
• Weighing range 1
• Weighing range 2
1 kg
2 kg
Graduations
• Weighing range 1
• Weighing range 2
50 g
100 g
Tare range
up to 300 kg
Accuracy following initial calibration
• Weighing range 1: 0 to 25 kg
• Weighing range 1: 25 to 100 kg
• Weighing range 1: 100 kg to 150 kg
• Weighing range 2: 0 to 50 kg
• Weighing range 2: 50 to 200 kg
• Weighing range 2: 200 kg to 300 kg
± 25 g
± 50 g
± 75 g
± 50 g
± 100 g
± 150 g