2 check content of calibration counter, Technical data, 1 general technical data – Seca 704 User Manual
Page 65: 2 check content of calibration, Counter
Technical data • 65
10.2 Check content of calibration counter
These seca scales have been calibrated. Calibration
may only be performed by authorised bodies. To
ensure this, the scales are equipped with a calibration
counter which registers every change in data relevant
to calibration.
If you wish to check that the scales are properly cali-
brated, proceed as follows:
1. Switch the scales off if necessary.
2. Hold down any key and start up the scales.
The current content of the calibration counter will
flash in the display for a few seconds.
3. Compare the output content of the calibration
counter with the figure on the calibration counter
sticker.
The two figures must match for calibration to be valid. If
the sticker and calibration counter do not match, recali-
bration is necessary. Contact your service partner or the
seca customer service department. Should recalibration
be necessary, the seca calibration counter sticker
shown above will be replaced by the adjacent recali-
bration sticker to identify the calibration counter read-
ing. This sticker will be secured with an additional seal
by a person authorised to perform recalibration. The
recalibration sticker is available from the seca customer
service department under the number 14-05-01-886.
11.TECHNICAL DATA
11.1 General technical data
Technical data for
seca 704
Dimensions
• Depth
• Width
• Height
520 mm
360 mm
930 mm
Weight
17 kg
Temperature range
+10° C to +40°C
Height of figures
25 mm
Power supply
Battery
Power supply unit (optional)