Cleaning, What to do if – Seca 704 User Manual
Page 62

62 •
6. Repeat steps 3. and 4. as appropriate for the
month (
), day (
dAy
), hour (
hour
) and minute
(
).
7. Confirm your selection each time.
After confirming your selection for Minute you will
leave the menu automatically.
These selections will be automatically transmitted
to the wireless printer.
The wireless printer automatically adds the date
and time to every printout.
NOTE:
For further operation of the wireless printer see
its user manual.
8. CLEANING
To clean the covering and housing use a household
detergent or standard disinfectant as required. Take
note of the instructions provided by the manufacturer.
Do not use scouring or caustic cleaning products,
spirit, petrol or similar substances for cleaning under
any circumstances. Such products may damage the
high-quality finishes.
9. WHAT TO DO IF …?
Malfunction
Cause/Remedy
… no weight is dis-
played during weighing?
The scales are not connected to the power supply
- Check the scales are switched on
- Check batteries are inserted
…
0.0
does not appear
before weighing?
A load was placed on the scales before being
switched on.
- Remove load
- Switch scales off and back on again
… a segment lights up
continuously or not at
all?
There is a fault at that point.
- Notify maintenance service.
…
is displayed?
Battery voltage is falling off.
- Insert new batteries as soon as possible
…
bAtt
is displayed?
Batteries are flat.
- Insert new batteries
…
St0p
is displayed?
The maximum load has been exceeded.
- Reduce load