The two figures must match for calibration to be valid. If
the sticker and calibration counter do not match, recali-
bration is necessary. Contact your service partner or the
seca customer service department. Once the scale has
been recalibrated, a new, updated calibration sticker is
used to identify the calibration counter state.This sticker
is attached with an extra seal issued by the person
authorized to perform the recalibration. The calibration
counter sticker can be ordered fromseca customer
service.

10.TECHNICAL DATA

Technical data for

seca 955

Dimensions

• Depth
• Width
• Height

956 mm
563 mm
897 mm

Weight

approx. 25 kg

Temperature range

+10° C to +40°C

Height of figures

25 mm

Power supply

Rechargeable battery block
Power pack

Power consumption

approx. 32 mA

Maximum runtime in rechargeable battery
operation

approx. 2.400 minutes

Medical device according to directive
93/42/EEC

Class I with measuring function

Calibration according to directive 2009/23/EC Class III

EN 60601-1:

• Class II totally insulated appliance:

• Type B electromedical device:

Maximum load

• Partial weighing range 1
• Partial weighing range 2

150 kg
250 kg

Minimum load

• Partial weighing range 1
• Partial weighing range 2

1.0 kg
2.0 kg