9. MAINTENANCE/RECALIBRATION

9.1

Information about maintenance and recalibration

We recommend having maintenance performed prior
to recalibration of the scales.

ATTENTION!

Incorrect measurement due to improper
maintenance

– Make sure that maintenance and repair are

only carried out by an authorised service
partner.

You can find your local service partner at

www.seca.com or just send an e-mail to
[email protected].

Ensure that recalibration is carried out by authorised
persons in accordance with the national legal require-
ments. The year in which initial calibration was per-
formed can be found behind the CE mark on the rating
plate above the number of the notified body 0109
(Office of Legal Metrology of the state of Hessen,
Germany).

Recalibration is required in all cases if one or more cali-
bration seals have been damaged or the content of the
calibration counter does not match the number on the
valid calibration counter sticker.

9.2

Check content of calibration counter

These seca scales have been calibrated. Calibration
may only be performed by authorised bodies. To
ensure this, the scales are equipped with a calibration
counter which registers every change in data relevant
to calibration.

If you wish to check that the scales are properly
calibrated, proceed as follows:

1. Switch the scales off if necessary.

2. Hold down any key and start up the scales.

The current content of the calibration counter will
flash in the display for a few seconds.

3. Compare the output content of the calibration

counter with the figure on the calibration counter
sticker.