What to do if …? • 27

English

8. WHAT TO DO IF …?

Malfunction

Cause/Remedy

… no weight is displayed
during weighing?

The scale has no power supply.

- Check whether the scale is switched on
- Check whether the rechargeable battery

block is inserted

0.00

does not appear

before weighing?

A load was placed on the scales before being
switched on.

- Remove load
- Switch scales off and back on again

… a segment lights up
continuously or not at
all?

There is a fault at that point.

- Notify maintenance service.

is displayed?

The voltage of the rechargeable battery block
drops.

- Charge the rechargeable battery block as

soon as possible

bAtt

is displayed?

Rechargeable battery block is empty.

- Charge rechargeable battery block

St0p

is displayed?

The maximum load has been exceeded.

- Reduce load

is displayed?

The ambient temperature of the scales is too high
or too low.

- Set up scales in an ambient temperature

between +10 °C and +40 °C

- Wait approx. 15 minutes until scales have

adjusted to ambient temperature

Er:X:11

is displayed?

The scales are too high, or too great a load is
applied at one point.

- Reduce load on scales or distribute weight-

more evenly

- Restart scales

Er:X:12

is displayed?

The scales have been switched on with too great
a load.

- Reduce load on scales
- Restart scales

Er:X:16

is displayed?

Oscillation of the scales has occurred, preventing
determination of the zero point.

- Restart scales